Phase Detection
Phase detection is the recommended technique for force gradient detection. Set the controls as follows:
- Channel 3 image Data Type: Phase
- Scan Line: Interleave
- Z range: 3 degrees
- Line Direction: Retrace
NOTE: It is important that the Line Direction be set to Retrace for both the main and interleave scans. If it is set to Trace, a band may appear along the left side of the images due to the time taken for the tip to move between the surface and the lift scan height.